GBS Test Pulse Rev. 3 Pulse Generator for Long-Term Insulation Tests
The test pulse design is designed for long-term insulation tests on test objects with capacitances up to 6 nF.
Converter for High-Voltage Test Transformer
· Internal sine wave generator: 50-200 Hz.
· Pulse width modulation of the sine wave from 16-24 kHz via full bridge.
· Converter voltage adjustable from 0-600 V, Imax=2.6 A, maximum power 1.3 kW
· Pulse current limit 110 A (cut-off threshold)
Pulse Transformer Control
· Control via full bridge
· Full-bridge power supply adjustable from 0-1000 V, 1.3 A max.
· Pulse current limit 100 A (cut-off threshold)
Pulse width 20-120 μs, nominal 94 μs, adjustable
· 2 selectable series inductors (150 μH, 300 μH) to adjust the pulse width and rise time to the load capacitance and voltage.
· Pulse repetition frequency (0.1 – 2.5 kHz), adjustable.
· Polarity: Adjustable positive, negative, or alternating depending on the polarity of the sine signal.
· Cosine-shaped pulse at the transformer output
Pulse transformer cascade
· Output transformers in a 5-way cascade circuit.
· Total transformation ratio selectable as 1:7.7, 1:10, or 1:15 via taps.
Maximum output voltage 15 kV with possible load capacitances of: * 6 nF with a pulse width of 73 μs at a
repetition frequency of 2.5 kHz * 10 nF with a pulse width of 103 μs at a repetition frequency of 1.25 kHz
· 1:1000 parallel voltage divider.
Gbjdpfxsu Egz Rj Agrji
· Overvoltage limitation at approx. 20-23 kV through parallel spark gaps.
· Thermistor for overtemperature detection in the first transformer stage.
· Ground-side current measurement socket 100 mV/A.
Breakdown detection, test
· If the maximum bridge currents for the transformer controls or
if the maximum currents of the bridge power supplies are exceeded, the high voltage is switched off.
· The device has a resettable operating hours counter to determine the test time.
Manufacturer's price approx. €59,500
Financing through our bank is also possible.
komplett-konzept.leasingo.de
Find more items - new and used - in our shop!
International shipping costs on request!
The test pulse design is designed for long-term insulation tests on test objects with capacitances up to 6 nF.
Converter for High-Voltage Test Transformer
· Internal sine wave generator: 50-200 Hz.
· Pulse width modulation of the sine wave from 16-24 kHz via full bridge.
· Converter voltage adjustable from 0-600 V, Imax=2.6 A, maximum power 1.3 kW
· Pulse current limit 110 A (cut-off threshold)
Pulse Transformer Control
· Control via full bridge
· Full-bridge power supply adjustable from 0-1000 V, 1.3 A max.
· Pulse current limit 100 A (cut-off threshold)
Pulse width 20-120 μs, nominal 94 μs, adjustable
· 2 selectable series inductors (150 μH, 300 μH) to adjust the pulse width and rise time to the load capacitance and voltage.
· Pulse repetition frequency (0.1 – 2.5 kHz), adjustable.
· Polarity: Adjustable positive, negative, or alternating depending on the polarity of the sine signal.
· Cosine-shaped pulse at the transformer output
Pulse transformer cascade
· Output transformers in a 5-way cascade circuit.
· Total transformation ratio selectable as 1:7.7, 1:10, or 1:15 via taps.
Maximum output voltage 15 kV with possible load capacitances of: * 6 nF with a pulse width of 73 μs at a
repetition frequency of 2.5 kHz * 10 nF with a pulse width of 103 μs at a repetition frequency of 1.25 kHz
· 1:1000 parallel voltage divider.
Gbjdpfxsu Egz Rj Agrji
· Overvoltage limitation at approx. 20-23 kV through parallel spark gaps.
· Thermistor for overtemperature detection in the first transformer stage.
· Ground-side current measurement socket 100 mV/A.
Breakdown detection, test
· If the maximum bridge currents for the transformer controls or
if the maximum currents of the bridge power supplies are exceeded, the high voltage is switched off.
· The device has a resettable operating hours counter to determine the test time.
Manufacturer's price approx. €59,500
Financing through our bank is also possible.
komplett-konzept.leasingo.de
Find more items - new and used - in our shop!
International shipping costs on request!
提供者
注意: 無料で登録するかログインして、すべての情報を取得してください。
登録日: 2008
590 オンライン広告
お問い合わせを送信
電話 & ファックス
+49 2862 ... 広告
これらの広告もご興味があるかもしれません。
小型広告
Soerendonk
9,258 km
ギアユニット/ギアボックス
Flender / SiemensK2S-2,5
Flender / SiemensK2S-2,5
小型広告
Wiefelstede
8,980 km
ベンチバイス 150mm
Heuer FrontSpannweite 235 mm
Heuer FrontSpannweite 235 mm
小型広告
ドイツ
9,058 km
変電所
Kalchschmid3076
Kalchschmid3076
小型広告
Tönisvorst
9,216 km
ベゼル 28427
mitlaufendSPH 150 mm
mitlaufendSPH 150 mm
あなたの掲載は正常に削除されました
エラーが発生しました