GBS Prüfpuls rev3 パルス発生器(長期絶縁試験用、最大6nF対応)GBS Elektronik
rev3 Isolationstests bis 6nF
GBS Prüfpuls rev3 パルス発生器(長期絶縁試験用、最大6nF対応)
GBS Elektronik
rev3 Isolationstests bis 6nF
VB 消費税別
€3,781
状態
中古
所在地
Borken 

基本データ
- メーカー:
- GBS Elektronik
- 状態:
- 中古
価格と所在地
VB 消費税別
€3,781
- 所在地:
- Dunkerstrasse 29, 46325 Borken/Burlo, Deutschland

電話する
オファーの詳細
- 広告ID:
- A17569468
- リファレンス番号:
- 21007
- 更新:
- 最終更新日:23.11.2026
説明
GBS Test Pulse rev3 Pulse Generator for Long-Term Insulation Tests up to 6 nF
The Test Pulse unit is designed for long-term insulation tests on test specimens with a capacitance of up to 6 nF.
Converter for High-Voltage Test Transformer
· Internal sine wave generator: 50-200 Hz.
· Pulse Width Modulation 16-24 kHz of the sine wave via a full bridge.
· Converter voltage adjustable from 0-600 V, Imax = 2.6 A, maximum power 1.3 kW.
· Pulse current limitation 110 A (shutdown threshold).
Pulse Transformer – Drive
· Drive via a full bridge.
· Full-bridge power supply adjustable from 0-1000 V, 1.3 A max.
· Pulse current limitation 100 A (shutdown threshold).
· Pulse width 20-120 μs, nominal 94 μs, adjustable.
· 2 selectable series inductances: 150 μH, 300 μH, to adjust the pulse width or rise times to the load capacitance and voltage.
· Pulse repetition frequency 0.1 – 2.5 kHz, adjustable.
· Polarity: Adjustable to positive, negative, or alternating, depending on the polarity of the sine wave.
· Cosine-shaped pulse at the transformer output.
Pulse Transformer Cascade
· Output transformers in a 5-stage cascaded circuit.
· Total turns ratio selectable as 1:7.7, 1:10, or 1:15 via taps.
· Maximum output voltage 15 kV with possible load capacitances of: * 6 nF with a pulse width of 73 μs at a
repetition frequency of 2.5 kHz * 10 nF with a pulse width of 103 μs at a repetition frequency of 1.25 kHz.
· Voltage divider 1:1000 in parallel.
· Overvoltage limitation at approximately 20-23 kV by parallel spark gaps.
· Thermistor for overtemperature detection in the 1st transformer stage.
· Earth-side current measurement socket 100 mV/A.
Breakdown Detection, Testing
· If the maximum bridge currents for the transformer drive or the maximum currents of the bridge power supplies are exceeded, the high voltage is switched off.
· The device has a resettable operating hour counter to determine the test time.
Manufacturer's price approximately €59,500.
Ndedpou Egz Rjfx Ahaod
You can find more items – new and used – in our shop!
International shipping costs on request!
この広告は自動翻訳されています。翻訳ミスがある可能性があります。
The Test Pulse unit is designed for long-term insulation tests on test specimens with a capacitance of up to 6 nF.
Converter for High-Voltage Test Transformer
· Internal sine wave generator: 50-200 Hz.
· Pulse Width Modulation 16-24 kHz of the sine wave via a full bridge.
· Converter voltage adjustable from 0-600 V, Imax = 2.6 A, maximum power 1.3 kW.
· Pulse current limitation 110 A (shutdown threshold).
Pulse Transformer – Drive
· Drive via a full bridge.
· Full-bridge power supply adjustable from 0-1000 V, 1.3 A max.
· Pulse current limitation 100 A (shutdown threshold).
· Pulse width 20-120 μs, nominal 94 μs, adjustable.
· 2 selectable series inductances: 150 μH, 300 μH, to adjust the pulse width or rise times to the load capacitance and voltage.
· Pulse repetition frequency 0.1 – 2.5 kHz, adjustable.
· Polarity: Adjustable to positive, negative, or alternating, depending on the polarity of the sine wave.
· Cosine-shaped pulse at the transformer output.
Pulse Transformer Cascade
· Output transformers in a 5-stage cascaded circuit.
· Total turns ratio selectable as 1:7.7, 1:10, or 1:15 via taps.
· Maximum output voltage 15 kV with possible load capacitances of: * 6 nF with a pulse width of 73 μs at a
repetition frequency of 2.5 kHz * 10 nF with a pulse width of 103 μs at a repetition frequency of 1.25 kHz.
· Voltage divider 1:1000 in parallel.
· Overvoltage limitation at approximately 20-23 kV by parallel spark gaps.
· Thermistor for overtemperature detection in the 1st transformer stage.
· Earth-side current measurement socket 100 mV/A.
Breakdown Detection, Testing
· If the maximum bridge currents for the transformer drive or the maximum currents of the bridge power supplies are exceeded, the high voltage is switched off.
· The device has a resettable operating hour counter to determine the test time.
Manufacturer's price approximately €59,500.
Ndedpou Egz Rjfx Ahaod
You can find more items – new and used – in our shop!
International shipping costs on request!
この広告は自動翻訳されています。翻訳ミスがある可能性があります。
提供者
注意: 無料で登録するかログインして、すべての情報を取得してください。
最後のオンライン:本日
登録日: 2008
535 オンライン広告
お問い合わせを送信
電話 & ファックス
+49 2862 ... 広告
これらの広告もご興味があるかもしれません。
成功
あなたのメッセージが送信されました。
小型広告
Wiefelstede
8,980 km
ベンチバイス 150mm
Heuer FrontSpannweite 235 mm
Heuer FrontSpannweite 235 mm
あなたの掲載は正常に削除されました
エラーが発生しました